International audience. The study of skin mechanical impedance is a clue to analyse its health. We use a Langevin Transducer at 60kHz to excite the skin. In return we observe the acoustic forces generated by the interaction. To achieve that, the system is identified and presented in the rotating reference frame, and we describe the design of two PI controllers that accurately control the vibration velocity. However, in vivo analysis can be tricky when it comes to safety and duration of tests. Thus, phantom tissues are used as substitute to simulate skin behaviour. Due to its mechanical properties similar to skin, and its simple manipulation, Dragon Skin ®, a silicone rubber type of phantom, was chosen. The paper presents the characterisation of the mechanical impedance of three silicon phantoms with different properties based on force observation. The results show that the phantoms can be indeed identified based on this method.

Consulter en ligne

Suggestions

Du même auteur

Low-frequency ultrasound skin mechanical impedance sensor for Pressure Ulcer diagnostic | Sales de Castro, Elis Marina

Low-frequency ultrasound skin mechanical impedance sensor for Pressure Ulce...

Archive ouverte: Communication dans un congrès

Sales de Castro, Elis Marina | 2022-06-15

National audience. Pressure ulcers are a concern not only for the patients who suffer from them, but also for public health. Its costs to the quality of life of patients and to the public finances can be reduced wit...

Low-frequency ultrasound device for pressure ulcer diagnosis | Sales de Castro, Elis Marina

Low-frequency ultrasound device for pressure ulcer diagnosis

Archive ouverte: Communication dans un congrès

Sales de Castro, Elis Marina | 2022-09-14

International audience

Low Frequency Ultrasound Diagnostics Sensor for Pressure Ulcers | Sales de Castro, Elis Marina

Low Frequency Ultrasound Diagnostics Sensor for Pressure Ulcers

Archive ouverte: Communication dans un congrès

Sales de Castro, Elis Marina | 2021-10

International audience. Mechanical properties of skin are a clue for the diagnostic of pressure ulcers (PU) [1]. This project proposes a low-frequency ultrasound (LFU) device to identify the mechanical impedance (MI...

Du même sujet

Low-frequency ultrasound skin mechanical impedance sensor for Pressure Ulcer diagnostic | Sales de Castro, Elis Marina

Low-frequency ultrasound skin mechanical impedance sensor for Pressure Ulce...

Archive ouverte: Communication dans un congrès

Sales de Castro, Elis Marina | 2022-06-15

National audience. Pressure ulcers are a concern not only for the patients who suffer from them, but also for public health. Its costs to the quality of life of patients and to the public finances can be reduced wit...

Modelling and control of a travelling wave in a finite beam, using multi-modal approach and vector control method | Ghenna, Sofiane

Modelling and control of a travelling wave in a finite beam, using multi-mo...

Archive ouverte: Communication dans un congrès

Ghenna, Sofiane | 2015-04-12

International audience. —This paper presents a new method to produce and control the vibration amplitude and direction of a travelling wave in a finite beam, using multi-modal approach. A closed loop control of the ...

LQR based MIMO-PID controller for the vector control of an underdamped harmonic oscillator | Kaci, Anis

LQR based MIMO-PID controller for the vector control of an underdamped harm...

Archive ouverte: Article de revue

Kaci, Anis | 2019-12-01

International audience. Modulated-Demodulated control (or vector control) allows to simultaneously impose amplitude and phase of a resonator. Moreover, the working frequency in the case of discrete-controller is sub...

Improving the PAD actuator performances by vector control | Giraud-Audine, Christophe

Improving the PAD actuator performances by vector control

Archive ouverte: Article de revue

Giraud-Audine, Christophe | 2018-12

International audience. In this work, we address the problem of pull-out and stalling of the vector control of the piezo-actuator drive (PAD). The model presented reveals the similarities with synchronous machine an...

Closed loop control of vibration field transient: Application to wave focusing | Kaci, Anis

Closed loop control of vibration field transient: Application to wave focus...

Archive ouverte: Article de revue

Kaci, Anis | 2022-03-01

In this article, several modes are controlled simultaneously both in phase and amplitude on an haptic display. To achieve this, modulation/demodulation control combined with mixed spatial/frequency filters is developed. It is then...

Modelling, identification and control of a Langevin transducer | Ghenna, Sofiane

Modelling, identification and control of a Langevin transducer

Archive ouverte: Communication dans un congrès

Ghenna, Sofiane | 2015-06-22

International audience. —The control of the vibration amplitude, and the resonance frequency tracking for ultrasonic transducer have been established. However, some applications require to control the vibration ampl...